Since its introduction in 1983, the Magnetic Property Measurement System (MPMS®) sample magnetometer has become the industry standard for performance and reliability. Utilizing Superconducting Quantum Interference Device (SQUID) technology, combined with patented enhancements, the MPMS-XL sample magnetometer achieves superior measurement sensitivity, dynamic range, and reproducibility otherwise unattainable.

The MPMS-XL provides solutions for a unique class of sensitive magnetic measurements in key areas such as high-temperature superconductivity, biochemistry, and magnetic recording media. Significant growth began in 1988 with the discovery of a new class of superconducting materials. While the basic application has not changed greatly, use has expanded to more than 530 installations worldwide. The MPMS product line provides the exceptional sensitivity of a SQUID-based magnetometer in a full featured, user friendly analytical instrument.

The modular MPMS design integrates a SQUID detection system, a precision temperature control unit residing in the bore of a high-field superconducting magnet, and a sophisticated computer operating system. Powerful software controls measurements, making data collection and analysis quick and easy. Easy-to-use Windows® software allows full automation of all system parameters.

Maximum Sample Size: 9 mm
Field Uniformity: 0.01% over 4 cm
Temperature Range: 1.9-400 K (800 K with optional oven)

MPMS EverCool™
A new dewar with an integrated helium reliquefication unit

The MPMS-XL Models offer high homogeneity magnet configurations - ± 1.0 Tesla, ± 5.0 Tesla and ± 7.0 Tesla. Each system includes:

  • Reciprocating Sample Measurement System for improved sensitivity and measurement speed
  • Continuous Low Temperature Control/Temperature Sweep with Enhanced Thermometry - Patented dual impedance design for continuous operation below 4.2 K, smooth transition through 4.2 K, and improved thermometry for Temperature Sweep Mode
  • MPMS MultiVu Windows-based Software Interface

MPMS® Options:
  • SQUID AC Susceptibility Measurement (*) - 0.1 Hz to 1KHz,
    sensitivity: 2 x l0-8 emu at 0 T
  • Ultra-Low Field Capability (*) ±0.05 G for the 5 T or 7T magnets
  • Reciprocating Sample Option (RSO) - DC Magnetization absolute
    sensitivity: 1 x 10-8 emu @ 2,500 Oe
  • Continuous Low Temperature Control/Temperature Sweep Mode (CLTC) -
    Sweep rate: 0.001 - 10 K/min.
  • Horizontal & Vertical Sample Rotators
  • Environmental Magnetic Shield
  • Sample Space Oven - increases temperature range to 800 K
  • Magnet Reset
  • Transverse Moment Detection System
  • Extended Dynamic Range (EDR) - expands the full-scale measuring capabilities to ± 300 emu.
  • External Device Control (EDC)
  • Fiber Optic Sample Holder
  • Low Field Profiling Option - includes 1 or 10 Gauss fluxgate
  • Standard & Nitrogen Jacketed Dewars
  • Manual Insertion Utility Probe
  • Resistivity Packages - (Voltmeter, Current Source, External
    Device Control, Manual Insertion Utility Probe

(*) Option not compatible with Transverse Moment Detection System